The Secondary Unit ID is a unique identification number that can be
configured into your ZENO
®
-3200 as the destination address for data output
messages.
The Secondary Unit ID can be set to any whole number within the range 0 to
9998.
The Secondary Unit ID cannot be the same as the Primary Unit ID.
73
To use the Secondary Unit ID, you also need to setup a Data Output Record (using the Data Output
Menu) with Field Type Code #5 in the appropriate location of your Data Output Message.
5.7. COMPASS SENSOR USAGE
If you are using a compass sensor as part of your sensor suite, two parameters are specified in the System
Functions Menu that affects its usage:
System Functions
Menu Line Item
No.
Parameter Name Valid Range
5 Add Compass To Vane 0 = NO or 1 = YES
6 Compass Offset -180 to 180 degrees
Line item #5, Add compass to vane, controls whether or not the ZENO
®
-3200 will add the measured
compass value to the measured wind direction (the vane) when executing a Wind Vector Average
Process
74
or Running Wind Vector Average Process.
75
If you are not measuring the wind direction, if
you have no compass, or if you are not processing the wind measurements using a Wind Vector Average
Process or Running Wind Vector Average Process, this option has no effect.
Line item #6, Compass offset, is a value that is added to the compass sensor measurement to compensate
for local variations in magnetic North. The value is added directly to the compass sensor reading
76
. If
you have no compass sensor in your sensor suite, this option has no effect.
5.8. BAROMETER ELEVATION SETTING
In the System Functions Menu, Line Item #7, the Barometer Elevation is used to correct the barometric
pressure reading to sea level, if your sensor suite includes a barometric pressure sensor and you are using
the Barometric Pressure Process.
77
The valid range of values for the Barometer Elevation line item is 0 to
5000 meters. If the Barometric Pressure Process
78
is not specified in your configuration, this line item has
no effect.
73
Refer to Section 5.5.
74
Refer to Appendix D, Category #2, Process #1.
75
Refer to Appendix D, Category #2, Process #2.
76
Refer to Section 10.3.4.
77
Refer to Appendix D, Category #2, Process #5.
78
Refer to Appendix D, Category #2, Process #5.
Kommentare zu diesen Handbüchern