ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page 71
Temperature Sensor is also used to know when the internal temperature falls below -40ºC. (Below -40ºC,
the ZENO
®
-3200 will not write Data Records to Flash Logging Memory because the circuitry may be
damaged.)
When you type T{ENTER} in the System Functions Menu, you will be prompted to enter the current
ambient air temperature. Ideally, you will be measuring the current ambient air temperature with an
accurate temperature transducer located very near your ZENO
®
-3200. If you need to move the ZENO
®
-
3200 to be near the transducer (or the transducer near the ZENO
®
-3200), it is recommended that you wait
15 minutes before entering the calibration temperature. This will allow either the Internal Temperature
Sensor or your external temperature transducer sufficient time to settle and provide an accurate reading.
When you enter the new temperature reading, you must include the temperature scale. The choices are
FAHRENHEIT (example input: 72.5F), CELSIUS (23.2C), or KELVIN (295.6K). The new
temperature that you type is automatically saved internally by the ZENO
®
-3200.
You do not need to save the configuration to EEPROM after entering the new temperature; ZENOSOFT
®
automatically saves the calibration settings to EEPROM automatically.
5.3. GETTING CURRENT ZENOSOFT
®
VERSION NUMBER
Option V in the System Functions Menu will display the version of ZENOSOFT
®
in use in your ZENO
®
-
3200.
If you need to call the Customer Service Department at Coastal
Environmental Systems, please note your version ZENOSOFT
®
before
calling.
5.4. CONTACT INFORMATION
Option I in the System Functions Menu will display information as to how to contact Coastal
Environmental Systems should the need arise.
5.5. ZENO
®
-3200 IDENTIFICATION NUMBER, THE PRIMARY UNIT ID
Many ZENO
®
-3200 users have more than one ZENO
®
-3200 collecting data in different locations, but
download the data from a central location. To identify the data coming from a particular ZENO
®
-3200, a
unique identification number can be set in the ZENO
®
-3200's configuration.