ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page 20
3.1.3. Understanding & Changing the Sample Interval and Power Usage
One of the most important aspects to understand about the ZENO
®
-3200 is timing. All data collection
done by the ZENO
®
-3200 is based upon regular timed intervals that are controlled by the ZENO
®
-3200's
internal clock. The main timed interval is called the Sample Interval.
The Sample Interval defines the overall, recurring period of time that all
data collection, processing, data logging and transmitting occurs.
Three timing values define the Sample Interval. They are located in the Sample Period Menu. Access to
the Sample Period Menu is done through the User Menu. From the User Menu, type S{ENTER} to enter
the Sample Period Menu. Your screen will look something like this:
> s
SAMPLE PERIOD MENU
(Cn/m) Change Item n To Value m (Q) Quit
(E) Save Parameters To EEPROM (H) Help
(U) User Menu
Item 1: 60 (Sample Interval Time)
Item 2: 55 (Sample Duration Time)
Item 3: 0 (Sample Time Offset)
>
The three values are defined as follows:
Sample Interval Time defines the length of each complete Sample Interval (the time between the
start of successive Sample Intervals)
Sample Duration Time defines the portion of the Sample Interval Time that the ZENO
®
-3200
actually performs all data collection, processing, data logging and transmitting
Sample Time Offset defines a time at the beginning of the Sample Interval to delay the start of the
actual Sample Duration Time
The sum of the Sample Duration Time and Sample Time Offset can be the
less than or equal to, but never exceed, the Sample Interval Time.
If you attempt to set the Sample Duration Time or the Sample Time Offset or a combination thereof to a
value greater than the Sample Interval Time, the ZENO
®
-3200 will respond with an error message and not
accept the requested change.
In our example "factory-installed" configuration illustrated above, the Sample Interval Time has been set
to 60 seconds and the Sample Duration Time has been set to 55 seconds. No Sample Time Offset is in
use. This is illustrated using the following timing diagram: