Campbell CM10K Spezifikationen Seite 50

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 95
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 49
50
Figure 7. Screenshot of the "Sensor Managment" home screen.
Sensor Calibration process Frequency
Temperature/RH Vaisala
HMP45C
Sent to manufacturer for calibration.
Fill in information on the Sensor Calibration Form
Submit information to the TEAM portal
Yearly
Precipitation (TB4) Level and pour 314 cc of water. Gauge should tip 39
times. If not send out for calibration
Fill in information on the Sensor Calibration Form
Submit information to the TEAM portal
Yearly
Solar Radiation (LI 200X)
Sent to manufacturer for calibration.
Fill in information on the Sensor Calibration Form
Submit information to the TEAM portal
Every two
years
1
Table 2. Summary of calibration process and schedule for the sensors used in the TEAM Climate Protocol.
1
Schedule for the first year of operation is slightly different; see Section 0 for details.
Seitenansicht 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 94 95

Kommentare zu diesen Handbüchern

Keine Kommentare