IRGASON Integrated CO
2
/H
2
O Open-Path Gas Analyzer and 3D Sonic Anemometer
16. Enter the known dew point (in C°) in the box labeled Span Dew Point and
press Span.
17. The zero-and-span procedure is now complete. Remove the shroud,
reconnect the IRGASON temperature probe, and prepare the site for
normal operation.
9.5 Replacing Desiccant and Scrubber Bottles
If more than two years have passed since replacing the desiccant / scrubber
bottles within the IRGASON, or if the zero-and-span readings have drifted
excessively (see Section 9.4, Zero and Span), the bottles should be replaced.
To do this, twist the scrubber-bottle covers of the upper and lower arms
counter-clockwise until they detach (they should loosen by hand). Remove the
IRGASON chemical bottles (p/n 26510) from inside the covers, and replace
them with new bottles lid-first (see Figure 9-5). Once this is done, firmly
screw the covers back on the arms. Allow the sensor to equilibrate for at least
24 hours (longer if in high humidity), and then perform a zero-and-span
procedure. If readings continue to be suspect, the sensor may need factory
recalibration (see Section 9.6, Factory Recalibration).
The scrubber bottles contain strong oxidizing agents.
Avoid direct contact with the chemicals inside the bottles.
Also, ensure your work area is well ventilated and free of
any reactive compounds, including liquid water. Store
used chemical bottles in a sealed container until disposal.
DANGER
The chemical bottles should be disposed of according to local and federal
regulations. For more information, MSDS (Material Safety Data Sheets) for
the chemicals are available in Appendix C and at www.campbellsci.com.
32