CS450/CS455 Pressure TransducerRevision: 1/13Copyright © 2009-2013Campbell Scientific, Inc.

Seite 2

CS450/CS455 Pressure Transducer 3.1 Connection Communicating with the CS450/CS455 requires the sensor to be either connected to a PC or an SDI-12 Rec

Seite 3 - Warranty

CS450/CS455 Pressure Transducer software package (e.g., LoggerNet) before and after the installation, or look in the Windows Device Manager list under

Seite 4 - Assistance

CS450/CS455 Pressure Transducer 3.1.2 Device Configuration Utility (version 1.12 or higher) The Device Configuration Utility (DevConfig) allows you t

Seite 5 - Table of Contents

CS450/CS455 Pressure Transducer FIGURE 3-3. Connect screen There are three settings that can be changed: SDI-12 address, Pressure/Level Units, and T

Seite 6

CS450/CS455 Pressure Transducer 3.1.3 SDI-12 Transparent Mode Transparent Mode allows direct communication with the CS450/CS455. This may require w

Seite 7 - 1. Introduction

CS450/CS455 Pressure Transducer CR10X-PB and CR510-PB dataloggers require “*#n” being entered at the prompt, where n is the control port being used. N

Seite 8 - 2. Specifications

CS450/CS455 Pressure Transducer FIGURE 3-6. HyperTerminal TABLE 3-4. RS-232 Terminal Commands Terminal Commands Values Returned 1 Serial Number,

Seite 9 - 3. Configuration

CS450/CS455 Pressure Transducer 3.2 SDI-12 Commands The CS450/CS455 uses an SDI-12-compatible hardware interface and supports a subset of the SDI-12

Seite 10 - 3.1 Connection

CS450/CS455 Pressure Transducer The CS450/CS455 is shipped from the factory with the address set to 0. The address on the CS450/CS455 can be changed b

Seite 11

CS450/CS455 Pressure Transducer Only SDI-12 instructions aM!, aM7!, and aM8! output the results obtained when using the multiplier and offset. The off

Seite 13

CS450/CS455 Pressure Transducer Pressure can be converted to feet of fresh water using the following simple equation: 1 psi = 2.31 feet of water For e

Seite 14

CS450/CS455 Pressure Transducer For example, if the correct elevation of the water, as measured by a staff gauge or other measurement device, is 2015.

Seite 15

CS450/CS455 Pressure Transducer 5.2 RS-232 TABLE 5-2. RS-232 Wiring Color CS450/CS455 Function Connection RS-232 9-pin Red +12vdc Power Source

Seite 16

CS450/CS455 Pressure Transducer Keyboard/Display users and SCWIN users can jump ahead to the Maintenance section. 6.2 Using CRBasic or Edlog 6.2.1 C

Seite 17 - 3.2 SDI-12 Commands

CS450/CS455 Pressure Transducer Sample Program for CR1000 Datalogger ‘CR1000 Series Datalogger ‘Declare the variable for the water level measurement

Seite 18

CS450/CS455 Pressure Transducer 3. In the Start Address field, type in the number of the first input location. 4. In the Number of InLocs field, typ

Seite 19 - 4. Installation

CS450/CS455 Pressure Transducer Periodic evaluation of the desiccant is vital for keeping the vent tube dry. The CS450/CS455 ships with the desiccant

Seite 20 - 4.6 Calculate Offset

CS450/CS455 Pressure Transducer Problem: Transducer appears to be operating properly but data shows a periodic or cyclic fluctuation not attributable

Seite 21 - 5. Wiring

CS450/CS455 Pressure Transducer 22

Seite 22 - 6. Programming

Appendix A. Calibration Certificate Each CS450/CS455 has been calibrated to meet printed accuracy specification at multiple temperature and pressure

Seite 23 - 6.2 Using CRBasic or Edlog

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Seite 24 - 6.2.2 Edlog Programming

Appendix A. Calibration Certificate A-2

Seite 25 - 7. Maintenance

Appendix A. Calibration Certificate bars = kPa / 100.0 psi = kPa * 0.145 FIGURE A-1. Calibration certificate A-3

Seite 26 - 8. Troubleshooting

Appendix A. Calibration Certificate A-4

Seite 28

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Seite 29

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Seite 30

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Seite 31

Table of Contents ii 7. Maintenance ...19 7.1 Every Visit...

Seite 32

CS450/CS455 Pressure Transducer The CS450/CS455 Submersible Pressure Transducer is designed for general pressure measurements. It uses the SDI-12 or

Seite 33

CS450/CS455 Pressure Transducer Care should be taken when opening the package not to damage or cut the cable jacket. If there is any question about da

Seite 34

CS450/CS455 Pressure Transducer Overpressure: 2x pressure range Dry Storage Temperature: -10° to 80°C Sensor will be damaged if it is encased in fro