TABLE 8-2. Multipliers and Offsets for Analog Outputs
Density
(mg m
-3
)
Voltage Output Multiplier
(mg m
-3
V
-1
)
Offset
(mg m
-3
)
CO
2
386.32 -102.59
H
2
O 8.65 -2.26
9. Maintenance
IRGASON operation requires six maintenance tasks:
routine maintenance
wick maintenance
analyzer window cleaning
zero and span
replacing the analyzer desiccant / scrubber bottles
factory recalibration
9.1 Routine Maintenance
Periodically do the following:
Check the humidity indicator card in the EC100 enclosure. If the highest
dot has turned pink, replace or recharge the desiccant bags. Replacement
desiccant may be purchased (p/n 6714) or old desiccant packs may be
recharged by heating in an oven. See the manual ENC10/12, ENC12/14,
ENC14/16, ENC16/18, available at www.campbellsci.com
, for more
details on recharging desiccant bags.
Make sure the LED Status Lights on the EC100 panel are green. If not,
verify that all connections are secure and that the instruments are powered.
Also check the individual diagnostic bits for the specific fault. See Table
10-2, Bits in the Sonic Diagnostic Flag and Table 10-3, Bits in the Gas
Diagnostic Flag.
9.2 Wick Maintenance
9.2.1 Sonic Wicks
The IRGASON cannot measure wind when water droplets completely obscure
the face of the sonic transducers. If the site experiences frequent rainfall, using
wicks on the sonic transducers may be desirable. The wicks help remove
water. Under some conditions, however, droplets may build up and remain on
the transducer faces. As soon as the water droplets evaporate or are removed,
the IRGASON will again measure wind. Remove the water droplets by
dabbing a cotton swab or tissue on the faces of the transducers.
The user must use discretion to determine if wicks are necessary. If wicks are
not used, gently remove the wicks from the sonic transducers, taking care not
to damage or peal the matching layer (rubber tips) from the brass housing of
the transducers.