Seitenansicht 20
Number 4250-2010
Revision 0.1
Date MAY 1996
Page 2 of 37
Coordinate with Handar for return, servicing, and recalibration of 540A/570A DCPs
that fail to operate within factory specifications.
Prepare purchase orders for factory servicing and recalibration of Handar 540A/570A
DCPs.
Replace the Handar 540A/570A internal battery as required.
Archive all DCP servicing records.
2.3 DATA COORDINATOR
The data coordinator shall:
Inform the instrument technician when a DCP is being removed from the field.
Provide the instrument technician with a description of the field problems observed
with the DCP.
3.0 REQUIRED EQUIPMENT AND MATERIALS
Specific instrumentation, tools, equipment, and materials required to service and test the
Handar 540A/570A DCP are as follows:
Calibrated voltage source - Datel Model DVC-350A or equivalent
RF wattmeter - Bird Model 43 with #250D power element and 50 ohm RF load or
equivalent
Digital voltmeter (4 1/2 digits)
Handar, Inc. Operating and Service Manual for 540A Multiple Access Data Acquisition
System, 560A Hydrologic Data Collection System, and 545A Programming Set
Handar, Inc. 570A Data Acquisition System Operating and Service Manual
Handar "TERM" program (DCP communication and interface software)
IBM PC-compatible computer
Spare circuit boards as required
Replacement internal 12 volt battery
Two (2) packs desiccant
Reference AT/RH sensor (Rotronics GT-L or equivalent)
Rotronics AT/RH Sensor (Model MP-100F, wired for use with the Handar 540A/570A
DCP)