Number 4250-2010
Revision 0.1
Date MAY 1996
Page 1 of 37
1.0 PURPOSE AND APPLICABILITY
This technical instruction (TI) describes procedures for servicing and calibration testing of
Handar 540A/570A Data Collection Platforms (DCPs). This TI, as referenced in Standard
Operating Procedure 4250, Servicing and Calibration of Optical Monitoring Dataloggers,
specifically describes procedures for:
Performing post-field inspections
Performing post-field timing and performance checks
Performing routine laboratory servicing and cleaning
Checking and performing laboratory modifications
Programming the DCP
Performing pre-field operational tests
Documenting all servicing tasks
Archiving servicing, repair, and calibration records
Handar 540A/570A DCPs are used as the primary dataloggers in the IMPROVE
transmissometer network.
2.0 RESPONSIBILITIES
2.1 PROJECT MANAGER
The project manager shall:
Verify that all Handar 540A/570A DCPs are serviced at least annually.
Verify that calibration, timing, and transmission checks are performed on all Handar
540A/570A DCPs at least annually.
Verify that all Handar 540A/570A DCPs are operating within factory specifications
prior to being shipped to the field for use at an operational monitoring site.
Verify that all Handar 540A/570A DCPs that do not operate within factory
specifications are returned to Handar for factory servicing and recalibration.
Ensure that all DCP servicing is documented and archived in accordance with the
procedures described in this TI.
2.2 INSTRUMENT TECHNICIAN
The instrument technician shall:
Perform and document all servicing, modifications, calibration checks, and operational
tests.