Number 4250-2000
Revision 1.0
Date MAY 1996
Page 5 of 7
After a few seconds delay, the memory check results will be
displayed. If all memory is installed and operating, the display will
read "11:111111." The eight (8) characters in the display represent
the eight (8) memory sockets numbered from left to right. A "1"
indicates a good chip is in the corresponding socket. A "0" indicates
the socket is empty or an error was detected in the chip. The five (5)
left-most characters of the display represent the 8K ram chips. The
three (3) right-most characters of the display are the 8K PROMs.
If the memory check results indicate that one or more memory chips
are faulty, return the instrument to Campbell Scientific for repair.
4.2 ANALOG INPUT CHECKS
CONNECT Connect the Datel voltage calibrator to the datalogger using the
VOLTAGE datalogger "analog inputs" test cable. This cable provides a
CALIBRATOR connection from the voltage output of the calibrator to each of the
eight (8) analog input channels of the datalogger.
DOWNLOAD Download the datalogger test program (21X_TEST.DLD) to the
TEST PROGRAM datalogger to be tested using the Campbell Scientific datalogger
communications software (SMCOM) and the Campbell Scientific
SC532 Peripheral Interface Module.
RUN TEST Press
*0
on the datalogger to compile and run the test program.
PROGRAM
SET VOLTAGES Set the calibrator to the input voltages specified on the Campbell
21X Datalogger Servicing Documentation Form (Figure 4-1). All
input voltages are specified in millivolts. All datalogger readings
should be recorded as millivolts.
RECORD Enter
*6
on the datalogger and record the datalogger display
DISPLAY reading (storage locations 01 - 08) for each of the eight analog
READINGS channels at each of the three input voltages specified on the
Campbell 21X Datalogger Servicing Documentation Form.
If the datalogger readings for any of the analog channels differ from
the specified values by more than ± 5.0 millivolts, return the
datalogger to Campbell Scientific for recalibration.
4.3 ANALOG OUTPUT CHECKS
The test program sets up a continuous DC voltage output on both analog output ports
(CAO 1 and CAO 2).
MEASURE Measure the output voltage at CAO ports 1 and 2 with a calibrated
OUTPUT and certified 4 ½ digit voltmeter. Record these measurements (in
VOLTAGE millivolts) on the Campbell 21X Datalogger Servicing