AM16/32B Relay Analog Multiplexer
Exposed ends of thermocouples measuring soil temperature should be
electrically insulated to prevent differences in ground potential among
the thermocouples from causing errors in the measured temperatures.
AM16/32B panel switch set to “4x16” mode.
Sensor to Multiplexer wiring - up to three thermocouples per SET; the
high side of each thermocouple is input into terminals ODD H,
ODD L, and EVEN H. The low sides of each thermocouple are
multiplexed in common through terminal EVEN L.
Multiplexer to Datalogger wiring - If the reference junction is at the
datalogger, then the wire that connects the COM ODD H, COM ODD
L, and COM EVEN H terminals to the datalogger should be of the
same composition as the high side of the thermocouples. Also, the
wire that connects COM EVEN L to datalogger ground should be of
the same composition as the low side of the thermocouples.
If the reference junction is at the AM16/32B (CSI 107 thermistor,
RTD, etc.), then copper wire should be used to connect COM
terminals to the datalogger.
6.6.3 Differential Thermocouple Measurement
AM16/32B panel switch set to “2x32” mode.
Multiplexer to Datalogger wiring - The wire types here can be
handled in one of two ways. If a reference junction (107 thermistor,
or RTD, etc.) is at the AM16/32B, then one copper wire may be run
between the COM terminals of the multiplexer and the datalogger
input channel.
If the reference junction is at the datalogger, then matching
thermocouple wire should be run between the COM terminals of the
multiplexer and the differential input channel on the datalogger
(observe TC wire polarity).
6.7 Mixed Sensor Types
In applications where sensor types are mixed, multiple hook-up
configurations and programming sequences are possible. Please
consult CSI for application assistance if you need to multiplex
markedly different sensor types in your application.
6.7.1 Mixed Sensor Example: Soil Moisture Blocks and
Thermocouples
AM16/32B panel switch set to “4x16” mode.
In this example, 16 thermocouples and 16 soil moisture blocks will be
multiplexed through the AM16/32B. One thermocouple and one soil
moisture block are input into each SET.