AM25T Solid State Multiplexer
Example 4. CR10(X) Program for Measuring 25 Type T Thermocouples with the Differential
Thermocouple Instruction and Long Lead Length Compensation on the RTD Measurement
*Table 1 Program
01: 10 Execution Interval (seconds)
01: Set Port(s) (P20) ;Configure Control Ports for 1 millisecond pulse
1: 9999 C8..C5 = nc/nc/nc/nc
2: 9933 C4..C1 = nc/nc/1ms/1ms
02: Do (P86)
1: 42 Set Port 2 High ;Turn On AM25T
03: Full Bridge (P6) ;Measure the output of the reference temp. full bridge, Vs
1: 1 Reps
2: 23 25 mV 60 Hz Rejection Range ;See Table 6-1 for older AM25T multiplexers
3: 1 DIFF Channel
4: 1 Excite all reps w/Exchan 1
5: 1200 mV Excitation ;See Table 6-1 for older AM25T multiplexers
6: 27 Loc [ Vs ]
7: 1 Mult
8: 0 Offset
04: Do (P86) ;Clock to advance to measure excitation
1: 71 Pulse Port 1
05: Full Bridge (P6) ;Measure the Excitation at the full bridge, Vx
1: 1 Reps
2: 25 2500 mV 60 Hz Rejection Range ;See Table 6-1 for older AM25T multiplexers
3: 1 DIFF Channel
4: 1 Excite all reps w/Exchan 1
5: 1200 mV Excitation ;See Table 6-1 for older AM25T multiplexers
6: 28 Loc [ Vx ]
7: .001 Mult
8: 0 Offset
06: Z=X/Y (P38) ;Calculate Vs/Vx
1: 27 X Loc [ Vs ]
2: 28 Y Loc [ Vx ]
3: 1 Z Loc [ RefTemp_C ]
07: Z=X*F (P37)
1: 1 X Loc [ RefTemp_C ]
2: -0.001 F
3: 1 Z Loc [ RefTemp_C ]
08: Z=X+F (P34)
1: 1 X Loc [ RefTemp_C ]
2: .09707 F
3: 1 Z Loc [ RefTemp_C ]