TDR100
20
Section 7 for discussion of the electrical conductivity measurement method and
Section 6.2.13.1 for discussion of Kp..
6.2.3 Parameter 3: Multiplexer and Probe Selection
Addressing scheme: ABCn
Where:
A -- Level 1 multiplexer channel
B -- Level 2 multiplexer channel
C -- Level 3 multiplexer channel
n -- the number of consecutive probes; addressing starts with the channel
specified for the highest level multiplexer.
Note: enter a 0 when a level is not used.
Example 1
An entry of 1008:
Entry
Description
1 channel 1 is specified for the level 1 multiplexer.
0 the level 2 multiplexer is not used.
0 the level 3 multiplexer is not used.
8 8 consecutive probes are read; the probes are connected to
channels 1, 2, 3, 4, 5, 6, 7, and 8 of the level 1 multiplexer.
Example 2
An entry of 3624:
Entry
Description
3 channel 3 is specified for the level 1 multiplexer.
6 channel 6 is specified for the level 2 multiplexer.
2 channel 2 is specified for the level 3 multiplexer.
4 4 consecutive probes are read; the probes are connected to
channels 2, 3, 4, and 5 of the level 3 multiplexer.
6.2.4 Parameter 4: Waveform Averaging
Sets the number of measurements averaged at a given distance from the
TDR100. A value of 4 is recommended. Higher values can be used when
noise is present. Averaging is useful when noise from power sources or when
noise of random nature is superimposed on the reflection waveform.
Averaging is accomplished by collecting n values at a given distance before
collecting values at the next distance increment where n is the value entered in
Average.