TDR100
30
9. Programming Examples
9.1 CR10X/CR23X Program Examples
Example 1, Measure and Record; Analog Measurements and Volumetric
Water Content
A CS605 or CS610 probe is connected directly to the TDR100. A cable length
of 6.25 meters was determined using PCTDR. The L
a
/L value is converted to
volumetric water content using the Ledieu , etal linear calibration function.
The program is executed once a minute. The array id, time, analog
measurements, and volumetric water content are output to Final Storage every
five minutes.
;{CR10X}
;
;Program Name = T100EX#1
;
*Table 1 Program
01: 60 Execution Interval (seconds)
;Measure the datalogger power supply voltage:
1: Batt Voltage (P10)
1: 1 Loc [ Bat_Volt ]
;Measure the internal temperature of the datalogger:
2: Internal Temperature (P17)
1: 2 Loc [ DL_Temp ]
;Set port 5 high to switch ON "Switched 12V" power from the datalogger to power the TDR100.
3: Do (P86)
1: 45 Set Port 5 High
;Measure the CS610 probe connected directly to the TDR100 and convert to water content:
4: TDR100 Measurement (P119)
1: 0 SDM Address
2: 0 La/L
3: 0000 MMMP Mux & Probe Selection
4: 4 Waveform Averaging
5: 1 Vp
6: 250 Points
7: 5.75 Cable Length (meters)
8: 5 Window Length (meters)
9: .3 Probe Length (meters)
10: 0.085 Probe Offset (meters)
11: 3 Loc [ WC_1 ]
12: .1138 Mult ; Multiplier for Liedu Calibration
13: -0.1758 Offset ; Offset for Liedu Calibration