AM25T Solid State Multiplexer
12
12: Beginning of Loop (P87) ;Loop through thermocouple channels
1: 0 Delay
2: 25 Loop Count
13: Do (P86) ;Clock the AM25T twice
1: 71 Pulse Port 1
14: Do (P86)
1: 71 Pulse Port 1
15: Thermocouple Temp (DIFF) (P14) ;Measure the connected thermocouple
1: 1 Reps
2: 21 2.5 mV 60 Hz Rejection Range
3: 1 DIFF Channel
4: 1 Type T (Copper-Constantan)
5: 1 Ref Temp (Deg. C) Loc [ RefTemp_C ]
6: 2-- Loc [ TC_1 ]
7: 1.0 Mult
8: 0.0 Offset
16: End (P95) ;End loop
17: Do (P86) ;Turn off AM25T
1: 52 Set Port 2 Low
6.2 Example 21X Programs
The two example 21X programs in this section cover measuring thermocouples
with the differential measurement instruction.
Input locations within the measurement loops must be indexed (--), see
Instruction 87 in the datalogger manual. To index a location, using Edlog ver
6.0 or greater, move the cursor to the Input location label and press “F4”. To
index a location with the datalogger keyboard, press the “C” key when
entering the location number and before the “A” key is pressed.
TABLE 6-3. Wiring for 21X Examples
Function AM25T 21X
+12V Power 12 V 12 V
Power and Shield Ground
Clock CLK C1
Reset RES C2
RTD Excitation EX E1
Analog Ground AG
Common High HI 1H
Common Low LO 1L