SDM-CVO4 Speci cations (Valid for a temperature range of -25° to +50°C, unless otherwise speci ed.
Power
Operating voltage: 12 Vdc nominal (8 to 16 V)
Current drain @ 12 Vdc:
Typical active quiescent current 27 or 54 mA,
depending on operating mode (no load on output
ports). To estimate the total current, add the
quiescent current to the sum of all output currents
multiplied by 1.5.
For example, if each port is at 10 mA output, the
total = 54 + (1.5•4•10) = 114 mA.
The CVO4 draws <0.5 mA with all outputs o .
Voltage Mode Output
Range: 0 to 10,000 mV
Resolution: 2.5 mV
Maximum output current: 30 mA per channel
Minimum load current: 5 μA if output < 200 mV
Accuracy @ +23°C: ±0.02% of set voltage + (±2.5 mV)
Accuracy @ -25° to +50°C: ±0.13% of set voltage + (±2.5 mV)
Current Mode Output
Range: 0 to 20 mA
Resolution: 5 μA
Minimum output current (leakage): 5 μA at +50°C
Accuracy @ +23°C: ±0.02% of full scale range + (±5 μA)
Accuracy @ -25° to +50°C:
Typical: ±0.1% of full scale range + (±5 μA)
Worst case: ±0.15% of full scale range + (±5 μA)
Minimum voltage drop across the internal current
regulating circuit: 2.5 V @ 20 mA current ow
Maximum input voltage relative to channel ground:
20 Vdc
Note: When in current mode, the V
o
terminal outputs an unregulated voltage
source at 15 V nominal (±10%), 30 mA maximum load.
Isolation
Description:
The SDM-CVO4 includes an internal isolation barrier
and components rated to provide signal isolation for
transients up to 1500 Vac (rms), 2500 Vdc nominal. The
isolation is between any output and the SDM-CVO4 ground
connection and between individual output channels.
Protection components are built-in, which will break down in a
controlled fashion at voltages close to this limit (see Operator’s
Manual Section 4.5, Safety Issues, for details).
Tested isolation:
Each channel of each unit is tested for isolation resistance at
500 Vdc. Pass level > 10 MΩ
Maximum recommended continuous operating voltage:
240 Vac rms di erential between an output and
datalogger ground, providing all issues relating to
local regulation for safe installation and operation
are followed (see Operator’s Manual Section 4.5,
Safety Issues)
Environmental
Operating temperature:
-25° to +50°C standard;
-40° to 80°C extended (available on special order)
Humidity: non-condensing
Physical
Dimensions:
17.6 x 11.1 x 2.4 cm (6.9” x 4.4” x 0.94”);
23.4 x 11.1 x 2. 4 cm (9.2” x 4.4” x 0.94”) when
tted with mounting brackets for enclosure.
The mounting bracket holes are spaced 20.3 cm (8”) apart
for attachment to an enclosure backplate.
Weight: 0.370 kg (13 oz.)
EMC Status:
Complies with EN55022-1:1998 and EN50082-1:1998
Power considerations
The SDM-CVO4 power requirements are large compared
to most Campbell Scienti c products - especially
when driving signi cant loads. Care must be taken
to ensure the power supply can deliver this higher
demand. Alkaline batteries are not recommended for
long-term applications.
The SDM-CVO4 has two internal power supplies - one
for channels 1 and 2 and one for channels 3 and 4. The
power supply for channels 3 and 4 is only turned on if
the datalogger sends an instruction that sets the output
of those channels. If channels 3 and 4 are not used,
the power consumption is approximately 20 mA lower
than when all outputs are used.
Where supported by the datalogger, and when the
application allows it, the SDM-CVO4 can be shut down
to reduce its power consumption to less than 0.5 mA.
In this state, all outputs are switched o .