DMM600 Duff Moisture Meter
3
FIGURE 1. DMM600
3.1 Sample Chamber
The sample chamber holds the sample during measurement. The sample
chamber is accessed by loosening the chamber access thumbscrew, rotating the
sample chamber cap slightly counter-clockwise and lifting the cap away from
the sample chamber.
A piston located inside the cap is attached to the compression knob by a
threaded shaft. With the sample in the sample chamber, the compression knob
is turned to compress the sample for measurement. The knob is also used to
relax the sample for removal.
The bottom of the sample chamber is a printed circuit board containing a
waveguide attached to piston. A high frequency signal is coupled to the
waveguide from the electronics in the main cylinder. The presence of water in
the sample determines the response of the circuit which is converted to water
content using a calibration equation. The waveguide is attached to a piston
which moves against precision springs to obtain correct compression of
sample.
The sample chamber can be removed for cleaning by removing the chamber
removal screw and rotating the sample chamber slightly counter-clockwise.
3.2 Electronics
The electronics are housed inside the main cylinder of the DMM600. The
electronics for the measurement are located below the sample chamber, and the
electronics for display and serial communications are located behind the
display.