Campbell CVF4 Spezifikationen Seite 35

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 60
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 34
CNR4 Net Radiometer
8.2 Testing the Pyrgeometer
It is assumed that the zero offset is no more than a few watts per square meter
(see second test in section 8.1).
The CNR4 body and the ambient air should be at the same temperature as
much as possible. Let the pyrgeometer rest for at least five minutes to regain
its thermal equilibrium. Set the voltmeter to its most sensitive range. To test if
the pyrgeometer is working properly, put your hand in front of the
pyrgeometer. The thermal radiation from your hand will cause the
pyrgeometer to generate a positive voltage when the surface temperature of
your hand is higher than the pyrgeometer temperature. The pyrgeometer will
generate a negative voltage if the hand is colder. The signal is proportional to
the temperature difference (see the rule of thumb in section 5.6). The radiation
that is emitted by the hand can be calculated by dividing the pyrgeometer
output by the sensor’s sensitivity value, and subsequently correcting for the
temperature, according to equation 5-2. If there are still no faults found, your
pyrgeometer is probably doing fine.
8.3 Testing the Thermistor
Using a multimeter, measure the resistance between the black and white wires
of the thermistor, and compare the value with the resistance values listed in
Table 5-1. The resistance should be around 10 k Ω at 25 °C, and the cable
resistance should add about 0.026 Ω per each foot of cable. When in doubt the
Pt-100 resistance (temperature) can be checked as well for reference.
8.4 Testing the Pt-100
Using a multimeter, measure the resistance between the two opposite wires of
the Pt-100 (gray-yellow, gray-brown, green-yellow, green-brown), and
compare the measured value with the resistance values listed in Table 5-2. The
resistance should be above 100 Ω (100 Ω at 0 °C), and the cable resistance
should add about 0.026 Ω per each foot of cable. When in doubt the thermistor
resistance (temperature) can be checked as well for reference.
9. Maintenance and Recalibration
The CNR4 is weatherproof, and is intended for a continuous outdoor use. The
materials used in pyranometer and pyrgeometer are robust and require very low
maintenance. For optimal results, however, proper care must be taken.
9.1 Cleaning Windows and Domes
The radiometer readings can be reduced if domes and windows are not clean.
The site operator should check the windows and domes of the CNR4 regularly,
and clean them as needed. Use distilled water or alcohol as cleaning solution,
being careful not to scratch the windows and domes during cleaning.
29
Seitenansicht 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 59 60

Kommentare zu diesen Handbüchern

Keine Kommentare