Model HFP01SC Self-Calibrating Soil Heat Flux Plate
Sensor: Thermopile and film heater
Heater Voltage Input: 9 to 15 Vdc
Heater Voltage Output: 0 to 2 Vdc
Expected Accuracy: ±3% of reading
Sensitivity (nominal): 50 μV W
–1
m
–2
Sensor Resistance (nominal): 2 Ω
Heater Resistance (nominal): 100 Ω
Duration of Calibration: ±3 min. @ 1.5 W; typically done every 3 to
6 hours
Weight without Cable: 200 g (7.05 oz)
6. Installation
6.1 Placement in Soil
The HFP01SC soil heat flux plates, the TCAV averaging soil temperature
probes, and the CS616, Water Content Reflectometer, are installed as shown in
FIGURE 6-1.
2 cm
6 cm
8 cm
Ground Surface
Up to 1 m
2.5 cm
Partial emplacement of the HFP01SC and the TCAV
sensors is shown for illustration purposes. All sensors
must be completely inserted into the soil face before
the hole is backfilled.
FIGURE 6-1. Placement of heat flux plates