
CS525 ISFET pH Probe
8
7. Troubleshooting
The most common causes for erroneous pH data include:
x poor sensor connections to the datalogger
x damaged cables
x scratched chip
Problem:
Output signal is at its maximum value.
Possible reason:
x Probe is not in fluid.
x Chip is polluted.
x Diaphragm is polluted.
x Chip is scratched.
Suggestion:
x Put probe in fluid.
x Clean probe using soapy water method.
x Probe can not be fixed if chip is scratched.
Problem:
Probe response is very slow.
Possible reason:
x Diaphragm is chipped or polluted
Suggestion:
x Clean probe with soapy water method.
Problem:
Probe signal is drifting.
Possible reason:
x Diaphragm can be dried out.
x Chip is scratched.
Suggestion:
x Soak probe for 10 minutes in saturated KC1 solution (CSI PN 16349).
x Probe can not be fixed.
Kommentare zu diesen Handbüchern