CS225 Temperature String
11
TABLE 7-3 SDI-12 Commands for the CS225
Serial number, location number,
depth value (in cm)
Serial number, location number, depth
value (in cm)
Read user resettable min temperature
Min. temperature - floating point (°C)
user resettable max t
Read lifetime min temperature
Min. temperature - floating point (°C)
Read lifetime max temperature
Max. temperature - floating point (°C)
reset user resettable min
Min. temperature - floating point (°C).
This value constitutes the minimum of all
1 second measurements taken since the
previous aR6! Command.
reset user resettable max
Max. temperature - floating point (°C).
This value constitutes the maximum of all
1 second measurements taken since the
previous aR6! Command.
in sequence are:
-9 / A-Z / a-z (no Address 0)
message with the
“{“
can correct units that
have conflicting addresses.
SDI-12 Identification command
X13CAMPBELLCS225 1.0 SN:XXXXX
7.2.1 SDI-12 Addressing
The CS225 comes pre-programmed with addresses from the factory. However,
if ever needed the address of temperature sensors can be changed. This may be
necessary if two CS225 strings need to be placed in the same SDI-12 channel
in order to avoid duplicate addresses on the same SDI-12 channel.
The starting address will be 1 and this will coincide with the first temperature
sensor, which is located at the end of the sensor array. It is recommended to
start the readdressing process with the largest temperature sensor address to
avoid duplicate addresses.
When readdressing temperature sensors you must avoid giving
multiple sensors the same address. If this does occur you will no
longer be able to communicate with these sensors. Use the
“aA{!” Command, where “a” is the affected address. This will
reset the affected sensors to their factory configured address
value.