CNR1 Net Radiometer
comparing the results of single-ended and differential measurements made
under the same conditions.
For the CR3000 and CR5000, one differential channel and a current excitation
channel are required to measure the PT-100.
For the other dataloggers, two differential channels and the 4WPB100 module
are required to measure the Pt-100 temperature sensor.
The 4WPB100 is included with the CNR1 sensor purchased from
CSI.
NOTE
FIGURE 6-1. 4WPB100 Module
6.1 Calibration Factor
Each CNR1 is provided with a ‘Certificate of Calibration’ by the manufacturer
that shows the sensor serial number and ‘sensitivity’, or calibration factor. The
serial number and sensitivity are also shown on a label attached to the sensor.
The calibration factor is in units of uV/(W m
-2
), which needs to be converted to
units of (W m
-2
)/mV for the multiplier parameter in the datalogger program.
To convert the units, divide the calibration factor into 1000. For example, if
the calibration factor is 7.30 uV/(W m
-2
), the multiplier is 1000/7.3 = 136.99
(W m
-2
)/mV.
6.2 Example Programs
6.2.1 Example 1, CR23X Program Using Differential Channels
Program Example 1 requires six differential channels and the 4WPB100
module to measure the four radiation outputs and the Pt-100 temperature
sensor. The program measures the sensors every 2 seconds and calculates and
stores the following data to final storage every 60 minutes:
11